Show Hamamatsu Avalanche Photo Diode 2505027969 - Archived data from Tue, 11. August 2020 17:29:57 CEST
This is all the information about APD 2505027969. If it is wrong, edit the data.
Subdetector specification: | |
---|---|
Serial: | 2505027969 |
Type: | Hamamatsu Avalanche Photo Diode |
Detector: | assigned to 2505027969/2310024965 |
Unit: | unassigned |
Preamp: | 0 |
Current location: | Bochum |
Installation information: | |
Label: | blue |
Manufacturer information: | |
Wafer position: | D12 |
Break-through voltage: | 413 V |
Voltage for Gain 100 (T=+25°C): | 384.5 V |
Dark current: | 4.5 nA |
Screening Logistics: | |
Available: | No |
Storage Box: | 177 |
Position in Box: | 26 |
EP1 batch: | 19 |
EP1 batch after irradiation: | none |
Shipment: | |
Grid number: | none |
Position in grid: | none |
Arrival for irradiation: | 02. Jul 2019 |
Sent for analysis after irradiation: | 09. Jul 2019 |
Return for assembly: | none |
Irradiation: | |
Date: | 04. Jul 2019 |
Dose used: | 37 Gy |
Temperature: | 20 °C |
Position: | none |
Bias voltage: | none |
Annealing: | |
Date: | 05. Jul 2019 |
Temperature: | 80 °C |
Duration: | 48 h |
Measurement results: | |
Voltage for Gain 100: | T = +20 °C: 384.9587277 V T = -25 °C: 349.0951381 V |
Voltage for Gain 150: | T = +20 °C: 392.7855004 V T = -25 °C: 356.5400814 V |
Voltage for Gain 200: | T = +20 °C: 397.1210472 V T = -25 °C: 360.7295793 V |
Gain/Voltage slope at M = 100: | T = +20 °C: 4.649002706 V-1 T = -25 °C: 4.911965945 V-1 |
Gain/Voltage slope at M = 150: | T = +20 °C: 9.356738837 V-1 T = -25 °C: 9.05147921 V-1 |
Gain/Voltage slope at M = 200: | T = +20 °C: 14.74160302 V-1 T = -25 °C: 16.08752499 V-1 |
Break-through voltage: | T = +20 °C: 412.221699 V T = -25 °C: 375.5090088 V |
Notes: |
Version history
Time | Author | Change comment |
---|---|---|
05. Feb 2021 11:11:07 CET | jreher | Updated box and position via API. |
11. Aug 2020 17:29:57 CEST | jreher | Updated APD characteristics via API. |
03. Apr 2020 10:08:40 CEST | jreher | Assigned to capsule by script |
13. Feb 2020 14:24:05 CET | jreher | Updated box and position via API. |
27. Nov 2019 20:00:16 CET | jreher | Updated APD location via API. |
25. Nov 2019 17:24:50 CET | jreher | Updated box and position via API. |
18. Jul 2019 10:26:48 CEST | spankonin | Updated APD batch assignment via API. |
09. Jul 2019 15:30:38 CEST | markus.moritz | |
05. Jul 2019 14:50:14 CEST | jreher | Updated APD location via API. |
24. Jun 2019 13:18:06 CEST | jgrochowski | Updated APD batch assignment via API. |
03. Jan 2019 00:12:09 CET | Tobias | Corrected dark current unit. |
12. Jan 2018 13:12:01 CET | Tobias | Imported from Hamamatsu datasheet. |