RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2504027881 - Archived data from Tue, 11. August 2020 17:29:42 CEST


This is all the information about APD 2504027881. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2504027881
Type: Hamamatsu Avalanche Photo Diode
Detector: assigned to 2504027881/2317025566
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: blue
   
Manufacturer information:  
Wafer position: H07
Break-through voltage: 408 V
Voltage for Gain 100 (T=+25°C): 379.9 V
Dark current: 4.1 nA
   
Screening Logistics:  
Available: No
Storage Box: 176
Position in Box: 1
EP1 batch: 17
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 02. Jul 2019
Sent for analysis after irradiation: 09. Jul 2019
Return for assembly: none
   
Irradiation:  
Date: 04. Jul 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 05. Jul 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 379.742557 V     T = -25 °C: 344.7320228 V
Voltage for Gain 150: T = +20 °C: 387.6261692 V     T = -25 °C: 352.1028682 V
Voltage for Gain 200: T = +20 °C: 392.0050423 V     T = -25 °C: 356.2390805 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.403501073 V-1     T = -25 °C: 4.70866557 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.520422148 V-1     T = -25 °C: 9.506531933 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.82123288 V-1     T = -25 °C: 15.27488988 V-1
Break-through voltage: T = +20 °C: 407.6050029 V     T = -25 °C: 371.4012498 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:10:57 CET jreher Updated box and position via API.
11. Aug 2020 17:29:42 CEST jreher Updated APD characteristics via API.
03. Apr 2020 09:46:08 CEST jreher Assigned to capsule by script
13. Feb 2020 14:23:54 CET jreher Updated box and position via API.
27. Nov 2019 20:00:03 CET jreher Updated APD location via API.
25. Nov 2019 17:24:44 CET jreher Updated box and position via API.
18. Jul 2019 10:27:55 CEST spankonin Updated APD batch assignment via API.
09. Jul 2019 13:02:56 CEST markus.moritz
05. Jul 2019 14:50:04 CEST jreher Updated APD location via API.
24. Jun 2019 13:05:54 CEST jgrochowski Updated APD batch assignment via API.
03. Jan 2019 00:12:07 CET Tobias Corrected dark current unit.
12. Jan 2018 13:11:59 CET Tobias Imported from Hamamatsu datasheet.