Subdetector specification:  
Serial: 2501027567
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: D13
Break-through voltage: 414 V
Voltage for Gain 100 (T=+25°C): 385.8 V
Dark current: 5.8 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 27
Position in Box: 29
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 16. Jan 2019
Sent for analysis after irradiation: 28. Jan 2019
Return for assembly: none
   
Irradiation:  
Date: 21. Jan 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 22. Jan 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 386.0473285 V     T = -25 °C: 350.4900842 V
Voltage for Gain 150: T = +20 °C: 393.8994766 V     T = -25 °C: 358.1350434 V
Voltage for Gain 200: T = +20 °C: 398.2507446 V     T = -25 °C: 362.4305552 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.535465652 V-1     T = -25 °C: 4.793305522 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.552598848 V-1     T = -25 °C: 9.092234873 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.03748112 V-1     T = -25 °C: 17.01494878 V-1
Break-through voltage: T = +20 °C: 413.455143 V     T = -25 °C: 377.0760074 V
   
Notes:


Version history

Time Author Change comment
05. May 2021 14:07:24 CEST oafedulidis Updated box and position via API.
05. Feb 2021 10:51:22 CET jreher Updated box and position via API.
03. Feb 2021 18:17:28 CET jreher Updated box and position via API.
11. Aug 2020 17:24:11 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:34:18 CET jreher Updated APD location via API.
25. Nov 2019 17:13:02 CET jreher Updated box and position via API.
04. Jul 2019 11:09:55 CEST jreher Updated APD location via API.
12. Jun 2019 16:47:27 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 16:32:22 CEST jreher Updated APD batch assignment via API.
28. Jan 2019 12:14:55 CET markus.moritz
03. Jan 2019 00:11:58 CET Tobias Corrected dark current unit.
12. Jan 2018 13:11:55 CET Tobias Imported from Hamamatsu datasheet.