RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2420027500 - Archived data from Wed, 21. April 2021 14:49:14 CEST


This is all the information about APD 2420027500. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2420027500
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: G13
Break-through voltage: 410 V
Voltage for Gain 100 (T=+25°C): 383.2 V
Dark current: 4.1 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 26
Position in Box: 30
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 16. Jan 2019
Sent for analysis after irradiation: 28. Jan 2019
Return for assembly: none
   
Irradiation:  
Date: 21. Jan 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 22. Jan 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 383.2958062 V     T = -25 °C: 347.5777241 V
Voltage for Gain 150: T = +20 °C: 391.1560079 V     T = -25 °C: 354.9842708 V
Voltage for Gain 200: T = +20 °C: 395.5026722 V     T = -25 °C: 359.1839814 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.384109633 V-1     T = -25 °C: 4.81318183 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.202316677 V-1     T = -25 °C: 9.373493523 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.75936454 V-1     T = -25 °C: 15.16439991 V-1
Break-through voltage: T = +20 °C: 409.9999367 V     T = -25 °C: 374.4414235 V
   
Notes:


Version history

Time Author Change comment
25. Sep 2021 16:11:32 CEST oafedulidis Assigned to capsule by script
10. May 2021 14:01:18 CEST oafedulidis Updated box and position via API.
21. Apr 2021 14:49:14 CEST oafedulidis Updated box and position via API.
05. Feb 2021 10:51:13 CET jreher Updated box and position via API.
03. Feb 2021 18:17:20 CET jreher Updated box and position via API.
11. Aug 2020 17:24:01 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:34:07 CET jreher Updated APD location via API.
25. Nov 2019 17:12:57 CET jreher Updated box and position via API.
04. Jul 2019 11:09:46 CEST jreher Updated APD location via API.
12. Jun 2019 16:47:23 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 16:32:18 CEST jreher Updated APD batch assignment via API.
28. Jan 2019 12:26:40 CET markus.moritz
03. Jan 2019 00:11:56 CET Tobias Corrected dark current unit.
12. Jan 2018 13:11:54 CET Tobias Imported from Hamamatsu datasheet.