Show Hamamatsu Avalanche Photo Diode 2317025583 - Archived data from Fri, 05. February 2021 11:18:20 CET
This is all the information about APD 2317025583. If it is wrong, edit the data.
Subdetector specification: | |
---|---|
Serial: | 2317025583 |
Type: | Hamamatsu Avalanche Photo Diode |
Detector: | unassigned |
Unit: | unassigned |
Preamp: | 0 |
Current location: | Bochum |
Installation information: | |
Label: | none |
Manufacturer information: | |
Wafer position: | B07 |
Break-through voltage: | 411 V |
Voltage for Gain 100 (T=+25°C): | 382.3 V |
Dark current: | 7.2 nA |
Screening Logistics: | |
Available: | Yes |
Storage Box: | 233 |
Position in Box: | 30 |
EP1 batch: | none |
EP1 batch after irradiation: | none |
Shipment: | |
Grid number: | none |
Position in grid: | none |
Arrival for irradiation: | 21. Aug 2019 |
Sent for analysis after irradiation: | 15. Oct 2019 |
Return for assembly: | none |
Irradiation: | |
Date: | 09. Oct 2019 |
Dose used: | 37 Gy |
Temperature: | 20 °C |
Position: | none |
Bias voltage: | none |
Annealing: | |
Date: | 12. Oct 2019 |
Temperature: | 80 °C |
Duration: | 48 h |
Measurement results: | |
Voltage for Gain 100: | T = +20 °C: 382.6363951 V T = -25 °C: 347.322815 V |
Voltage for Gain 150: | T = +20 °C: 390.467346 V T = -25 °C: 354.726686 V |
Voltage for Gain 200: | T = +20 °C: 394.8165825 V T = -25 °C: 358.8667417 V |
Gain/Voltage slope at M = 100: | T = +20 °C: 4.430274677 V-1 T = -25 °C: 4.7680913 V-1 |
Gain/Voltage slope at M = 150: | T = +20 °C: 8.58892072 V-1 T = -25 °C: 9.778157901 V-1 |
Gain/Voltage slope at M = 200: | T = +20 °C: 14.91637721 V-1 T = -25 °C: 15.6241608 V-1 |
Break-through voltage: | T = +20 °C: 410.3131785 V T = -25 °C: 374.2159189 V |
Notes: |
Version history
Time | Author | Change comment |
---|---|---|
15. Nov 2022 12:54:33 CET | oafedulidis | Assigned to capsule by script |
15. Apr 2021 17:00:46 CEST | jreher | Updated box and position via API. |
05. Feb 2021 11:18:20 CET | jreher | Updated box and position via API. |
11. Aug 2020 17:41:54 CEST | jreher | Updated APD characteristics via API. |
21. Jul 2020 14:29:49 CEST | jreher | Updated box and position via API. |
16. Jan 2020 12:05:55 CET | jreher | Updated date sent for analysis after screening via API. |
16. Jan 2020 12:04:05 CET | jreher | Updated annealing info via API. |
16. Jan 2020 12:01:16 CET | jreher | Updated irradiation info via API. |
16. Jan 2020 11:58:20 CET | jreher | Updated arrival for irradiation via API. |
16. Jan 2020 11:57:22 CET | jreher | Updated arrival for irradiation via API. |
27. Nov 2019 20:10:17 CET | jreher | Updated APD location via API. |
25. Nov 2019 17:28:59 CET | jreher | Updated box and position via API. |
22. Oct 2019 13:08:17 CEST | spankonin | Updated APD location via API. |
18. Oct 2019 14:12:39 CEST | jgrochowski | Updated APD batch assignment via API. |
23. Jun 2016 12:00:11 CEST | Tobias | Imported from Hamamatsu datasheet. |