RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2317025558 - Archived data from Fri, 27. August 2021 11:14:28 CEST


This is all the information about APD 2317025558. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2317025558
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: C10
Break-through voltage: 415 V
Voltage for Gain 100 (T=+25°C): 386.3 V
Dark current: 6.3 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 225
Position in Box: 45
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 21. Aug 2019
Sent for analysis after irradiation: 15. Oct 2019
Return for assembly: none
   
Irradiation:  
Date: 09. Oct 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 12. Oct 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 386.7895827 V     T = -25 °C: 350.9546486 V
Voltage for Gain 150: T = +20 °C: 394.6022365 V     T = -25 °C: 358.3527029 V
Voltage for Gain 200: T = +20 °C: 398.9208818 V     T = -25 °C: 362.4987035 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.672524329 V-1     T = -25 °C: 4.904836563 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.38632697 V-1     T = -25 °C: 9.098787129 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.74866447 V-1     T = -25 °C: 16.2786621 V-1
Break-through voltage: T = +20 °C: 414.308316 V     T = -25 °C: 377.7826829 V
   
Notes:


Version history

Time Author Change comment
10. Jun 2022 12:31:06 CEST oafedulidis Assigned to capsule by script
27. Aug 2021 11:14:28 CEST oafedulidis Updated box and position via API.
05. Feb 2021 11:17:25 CET jreher Updated box and position via API.
11. Aug 2020 17:41:48 CEST jreher Updated APD characteristics via API.
21. Jul 2020 14:28:36 CEST jreher Updated box and position via API.
16. Jan 2020 12:05:53 CET jreher Updated date sent for analysis after screening via API.
16. Jan 2020 12:04:04 CET jreher Updated annealing info via API.
16. Jan 2020 12:01:14 CET jreher Updated irradiation info via API.
16. Jan 2020 11:58:17 CET jreher Updated arrival for irradiation via API.
16. Jan 2020 11:57:20 CET jreher Updated arrival for irradiation via API.
27. Nov 2019 20:08:54 CET jreher Updated APD location via API.
25. Nov 2019 17:28:28 CET jreher Updated box and position via API.
22. Oct 2019 12:57:29 CEST spankonin Updated APD location via API.
18. Oct 2019 14:11:39 CEST jgrochowski Updated APD batch assignment via API.
23. Jun 2016 12:00:10 CEST Tobias Imported from Hamamatsu datasheet.