Show Hamamatsu Avalanche Photo Diode 2316025516 - Archived data from Tue, 11. August 2020 17:42:30 CEST
This is all the information about APD 2316025516. If it is wrong, edit the data.
Subdetector specification: | |
---|---|
Serial: | 2316025516 |
Type: | Hamamatsu Avalanche Photo Diode |
Detector: | unassigned |
Unit: | unassigned |
Preamp: | 0 |
Current location: | Bochum |
Installation information: | |
Label: | none |
Manufacturer information: | |
Wafer position: | C14 |
Break-through voltage: | 419 V |
Voltage for Gain 100 (T=+25°C): | 390.3 V |
Dark current: | 5.3 nA |
Screening Logistics: | |
Available: | Yes |
Storage Box: | 22 |
Position in Box: | 50 |
EP1 batch: | none |
EP1 batch after irradiation: | none |
Shipment: | |
Grid number: | none |
Position in grid: | none |
Arrival for irradiation: | 21. Aug 2019 |
Sent for analysis after irradiation: | 15. Oct 2019 |
Return for assembly: | none |
Irradiation: | |
Date: | 09. Oct 2019 |
Dose used: | 37 Gy |
Temperature: | 20 °C |
Position: | none |
Bias voltage: | none |
Annealing: | |
Date: | 12. Oct 2019 |
Temperature: | 80 °C |
Duration: | 48 h |
Measurement results: | |
Voltage for Gain 100: | T = +20 °C: 390.6174809 V T = -25 °C: 354.5876553 V |
Voltage for Gain 150: | T = +20 °C: 398.4843173 V T = -25 °C: 362.1550323 V |
Voltage for Gain 200: | T = +20 °C: 402.8357085 V T = -25 °C: 366.4152681 V |
Gain/Voltage slope at M = 100: | T = +20 °C: 4.413442189 V-1 T = -25 °C: 4.578306405 V-1 |
Gain/Voltage slope at M = 150: | T = +20 °C: 8.582173631 V-1 T = -25 °C: 9.103277447 V-1 |
Gain/Voltage slope at M = 200: | T = +20 °C: 14.85976425 V-1 T = -25 °C: 14.22878076 V-1 |
Break-through voltage: | T = +20 °C: 417.7246084 V T = -25 °C: 381.7696136 V |
Notes: |
Version history
Time | Author | Change comment |
---|---|---|
26. Apr 2021 14:29:57 CEST | oafedulidis | Updated box and position via API. |
05. Feb 2021 11:17:34 CET | jreher | Updated box and position via API. |
11. Aug 2020 17:42:30 CEST | jreher | Updated APD characteristics via API. |
21. Jul 2020 14:28:48 CEST | jreher | Updated box and position via API. |
16. Jan 2020 12:05:50 CET | jreher | Updated date sent for analysis after screening via API. |
16. Jan 2020 12:04:01 CET | jreher | Updated annealing info via API. |
16. Jan 2020 12:01:11 CET | jreher | Updated irradiation info via API. |
16. Jan 2020 11:58:12 CET | jreher | Updated arrival for irradiation via API. |
16. Jan 2020 11:57:17 CET | jreher | Updated arrival for irradiation via API. |
27. Nov 2019 20:09:06 CET | jreher | Updated APD location via API. |
25. Nov 2019 17:28:32 CET | jreher | Updated box and position via API. |
22. Oct 2019 16:25:56 CEST | spankonin | Updated APD batch assignment via API. |
22. Oct 2019 16:10:08 CEST | spankonin | Updated APD location via API. |
23. Jun 2016 12:00:09 CEST | Tobias | Imported from Hamamatsu datasheet. |