Show Hamamatsu Avalanche Photo Diode 2314025317 - Archived data from Mon, 17. May 2021 09:14:53 CEST
This is all the information about APD 2314025317. If it is wrong, edit the data.
Subdetector specification: | |
---|---|
Serial: | 2314025317 |
Type: | Hamamatsu Avalanche Photo Diode |
Detector: | unassigned |
Unit: | unassigned |
Preamp: | 0 |
Current location: | Bochum |
Installation information: | |
Label: | none |
Manufacturer information: | |
Wafer position: | H08 |
Break-through voltage: | 410 V |
Voltage for Gain 100 (T=+25°C): | 382.3 V |
Dark current: | 8.4 nA |
Screening Logistics: | |
Available: | Yes |
Storage Box: | 23 |
Position in Box: | 44 |
EP1 batch: | none |
EP1 batch after irradiation: | none |
Shipment: | |
Grid number: | none |
Position in grid: | none |
Arrival for irradiation: | 21. Aug 2019 |
Sent for analysis after irradiation: | 15. Oct 2019 |
Return for assembly: | none |
Irradiation: | |
Date: | 09. Oct 2019 |
Dose used: | 37 Gy |
Temperature: | 20 °C |
Position: | none |
Bias voltage: | none |
Annealing: | |
Date: | 12. Oct 2019 |
Temperature: | 80 °C |
Duration: | 48 h |
Measurement results: | |
Voltage for Gain 100: | T = +20 °C: 381.7634443 V T = -25 °C: 346.6915773 V |
Voltage for Gain 150: | T = +20 °C: 389.5750577 V T = -25 °C: 354.0728849 V |
Voltage for Gain 200: | T = +20 °C: 393.9157927 V T = -25 °C: 358.2206116 V |
Gain/Voltage slope at M = 100: | T = +20 °C: 4.738234355 V-1 T = -25 °C: 4.709550609 V-1 |
Gain/Voltage slope at M = 150: | T = +20 °C: 8.561978198 V-1 T = -25 °C: 9.474351667 V-1 |
Gain/Voltage slope at M = 200: | T = +20 °C: 14.82687982 V-1 T = -25 °C: 14.90840996 V-1 |
Break-through voltage: | T = +20 °C: 404.4626609 V T = -25 °C: 373.2103051 V |
Notes: |
Version history
Time | Author | Change comment |
---|---|---|
11. Jul 2022 14:24:12 CEST | oafedulidis | Assigned to capsule by script |
17. May 2021 09:14:53 CEST | oafedulidis | Updated box and position via API. |
26. Apr 2021 14:28:04 CEST | oafedulidis | Updated box and position via API. |
05. Feb 2021 11:16:10 CET | jreher | Updated box and position via API. |
11. Aug 2020 17:42:06 CEST | jreher | Updated APD characteristics via API. |
21. Jul 2020 14:26:58 CEST | jreher | Updated box and position via API. |
16. Jan 2020 12:05:41 CET | jreher | Updated date sent for analysis after screening via API. |
16. Jan 2020 12:03:51 CET | jreher | Updated annealing info via API. |
16. Jan 2020 12:01:01 CET | jreher | Updated irradiation info via API. |
16. Jan 2020 11:57:54 CET | jreher | Updated arrival for irradiation via API. |
16. Jan 2020 11:57:06 CET | jreher | Updated arrival for irradiation via API. |
27. Nov 2019 20:07:12 CET | jreher | Updated APD location via API. |
25. Nov 2019 17:27:44 CET | jreher | Updated box and position via API. |
28. Oct 2019 11:13:08 CET | jreher | Updated APD location via API. |
22. Oct 2019 16:22:41 CEST | spankonin | Updated APD batch assignment via API. |
23. Jun 2016 11:59:12 CEST | Tobias | Imported from Hamamatsu datasheet. |