Show Hamamatsu Avalanche Photo Diode 2003021691 - Archived data from Fri, 05. February 2021 11:20:46 CET
This is all the information about APD 2003021691. If it is wrong, edit the data.
Subdetector specification: | |
---|---|
Serial: | 2003021691 |
Type: | Hamamatsu Avalanche Photo Diode |
Detector: | unassigned |
Unit: | unassigned |
Preamp: | 0 |
Current location: | Bochum |
Installation information: | |
Label: | none |
Manufacturer information: | |
Wafer position: | G07 |
Break-through voltage: | 407 V |
Voltage for Gain 100 (T=+25°C): | 378.1 V |
Dark current: | 6.2 nA |
Screening Logistics: | |
Available: | Yes |
Storage Box: | 252 |
Position in Box: | 23 |
EP1 batch: | none |
EP1 batch after irradiation: | none |
Shipment: | |
Grid number: | none |
Position in grid: | none |
Arrival for irradiation: | 21. Aug 2019 |
Sent for analysis after irradiation: | 15. Oct 2019 |
Return for assembly: | none |
Irradiation: | |
Date: | 09. Oct 2019 |
Dose used: | 37 Gy |
Temperature: | 20 °C |
Position: | none |
Bias voltage: | none |
Annealing: | |
Date: | 12. Oct 2019 |
Temperature: | 80 °C |
Duration: | 48 h |
Measurement results: | |
Voltage for Gain 100: | T = +20 °C: 378.8007889 V T = -25 °C: 343.6370012 V |
Voltage for Gain 150: | T = +20 °C: 386.7115929 V T = -25 °C: 351.0512835 V |
Voltage for Gain 200: | T = +20 °C: 391.1177876 V T = -25 °C: 355.2379415 V |
Gain/Voltage slope at M = 100: | T = +20 °C: 4.628321236 V-1 T = -25 °C: 4.662614426 V-1 |
Gain/Voltage slope at M = 150: | T = +20 °C: 9.106398687 V-1 T = -25 °C: 9.300404284 V-1 |
Gain/Voltage slope at M = 200: | T = +20 °C: 14.07358557 V-1 T = -25 °C: 14.56613139 V-1 |
Break-through voltage: | T = +20 °C: 406.4370329 V T = -25 °C: 370.6211895 V |
Notes: |
Version history
Time | Author | Change comment |
---|---|---|
15. Nov 2022 12:53:47 CET | oafedulidis | Assigned to capsule by script |
16. Dec 2021 13:41:49 CET | oafedulidis | Updated box and position via API. |
21. Apr 2021 14:52:56 CEST | oafedulidis | Updated box and position via API. |
05. Feb 2021 11:20:46 CET | jreher | Updated box and position via API. |
11. Aug 2020 17:42:48 CEST | jreher | Updated APD characteristics via API. |
21. Jul 2020 14:32:55 CEST | jreher | Updated box and position via API. |
16. Jan 2020 12:06:15 CET | jreher | Updated date sent for analysis after screening via API. |
16. Jan 2020 12:04:25 CET | jreher | Updated annealing info via API. |
16. Jan 2020 12:01:35 CET | jreher | Updated irradiation info via API. |
16. Jan 2020 11:58:43 CET | jreher | Updated arrival for irradiation via API. |
16. Jan 2020 11:57:42 CET | jreher | Updated arrival for irradiation via API. |
27. Nov 2019 20:13:28 CET | jreher | Updated APD location via API. |
25. Nov 2019 17:30:04 CET | jreher | Updated box and position via API. |
22. Oct 2019 16:34:30 CEST | spankonin | Updated APD batch assignment via API. |
22. Oct 2019 16:15:26 CEST | spankonin | Updated APD location via API. |
23. Jun 2016 11:52:42 CEST | Tobias | Imported from Hamamatsu datasheet. |