RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 1914021360 - Archived data from Tue, 11. August 2020 17:47:07 CEST


This is all the information about APD 1914021360. If it is wrong, edit the data.

Subdetector specification:  
Serial: 1914021360
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: G10
Break-through voltage: 410 V
Voltage for Gain 100 (T=+25°C): 381.3 V
Dark current: 17.2 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 208
Position in Box: 35
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 382.0333189 V     T = -25 °C: 346.5363144 V
Voltage for Gain 150: T = +20 °C: 389.8813641 V     T = -25 °C: 353.9780467 V
Voltage for Gain 200: T = +20 °C: 394.2436186 V     T = -25 °C: 358.1653994 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.579467196 V-1     T = -25 °C: 4.669544624 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.041762102 V-1     T = -25 °C: 9.374096666 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.10247612 V-1     T = -25 °C: 14.70713485 V-1
Break-through voltage: T = +20 °C: 409.5165487 V     T = -25 °C: 373.4723313 V
   
Notes:


Version history

Time Author Change comment
15. Apr 2021 16:17:55 CEST oafedulidis Assigned to capsule by script
16. Feb 2021 12:28:23 CET jreher Updated box and position via API.
05. Feb 2021 11:15:09 CET jreher Updated box and position via API.
11. Aug 2020 17:47:07 CEST jreher Updated APD characteristics via API.
27. Nov 2019 20:05:54 CET jreher Updated APD location via API.
25. Nov 2019 17:27:13 CET jreher Updated box and position via API.
18. Nov 2019 18:19:45 CET spankonin Updated APD location via API.
18. Nov 2019 18:16:48 CET spankonin Updated APD batch assignment via API.
23. Jun 2016 11:50:00 CEST Tobias Imported from Hamamatsu datasheet.