RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 1913021278 - Archived data from Tue, 11. August 2020 17:49:40 CEST


This is all the information about APD 1913021278. If it is wrong, edit the data.

Subdetector specification:  
Serial: 1913021278
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: D13
Break-through voltage: 413 V
Voltage for Gain 100 (T=+25°C): 385 V
Dark current: 13.1 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 265
Position in Box: 6
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 385.355068 V     T = -25 °C: 349.7001263 V
Voltage for Gain 150: T = +20 °C: 393.1336817 V     T = -25 °C: 357.1767729 V
Voltage for Gain 200: T = +20 °C: 397.4580114 V     T = -25 °C: 361.3916091 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.48373539 V-1     T = -25 °C: 4.96608512 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.900178941 V-1     T = -25 °C: 9.156725406 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.64132226 V-1     T = -25 °C: 16.16150013 V-1
Break-through voltage: T = +20 °C: 412.5872319 V     T = -25 °C: 376.5893083 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:22:25 CET jreher Updated box and position via API.
07. Oct 2020 15:44:28 CEST jreher Updated APD location via API.
11. Aug 2020 17:49:40 CEST jreher Updated APD characteristics via API.
21. Jul 2020 14:35:13 CEST jreher Updated box and position via API.
27. Nov 2019 20:15:27 CET jreher Updated APD location via API.
25. Nov 2019 17:30:41 CET jreher Updated box and position via API.
25. Nov 2019 17:17:54 CET spankonin Updated APD batch assignment via API.
23. Jun 2016 11:49:58 CEST Tobias Imported from Hamamatsu datasheet.