RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 1906020749 - Archived data from Fri, 05. February 2021 11:13:21 CET


This is all the information about APD 1906020749. If it is wrong, edit the data.

Subdetector specification:  
Serial: 1906020749
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: D13
Break-through voltage: 407 V
Voltage for Gain 100 (T=+25°C): 378.4 V
Dark current: 23.9 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 194
Position in Box: 3
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 18. Apr 2019
Sent for analysis after irradiation: 29. Jul 2019
Return for assembly: none
   
Irradiation:  
Date: 25. Jul 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 27. Jul 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 378.9980747 V     T = -25 °C: 343.590336 V
Voltage for Gain 150: T = +20 °C: 386.8521505 V     T = -25 °C: 350.9214016 V
Voltage for Gain 200: T = +20 °C: 391.2341722 V     T = -25 °C: 355.0502425 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.575212657 V-1     T = -25 °C: 4.720596626 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.001762726 V-1     T = -25 °C: 9.480475705 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 13.87059722 V-1     T = -25 °C: 15.03586289 V-1
Break-through voltage: T = +20 °C: 407.0265039 V     T = -25 °C: 370.4761146 V
   
Notes:


Version history

Time Author Change comment
25. Oct 2021 14:01:48 CEST oafedulidis Assigned to capsule by script
15. Apr 2021 17:01:18 CEST jreher Updated box and position via API.
05. Feb 2021 11:13:21 CET jreher Updated box and position via API.
11. Aug 2020 17:35:27 CEST jreher Updated APD characteristics via API.
27. Nov 2019 20:03:22 CET jreher Updated APD location via API.
25. Nov 2019 17:26:08 CET jreher Updated box and position via API.
16. Aug 2019 12:20:56 CEST jreher Updated APD location via API.
09. Aug 2019 16:47:35 CEST jreher Updated APD batch assignment via API.
09. Aug 2019 16:41:27 CEST jreher Updated APD batch assignment via API.
09. Aug 2019 16:35:28 CEST jreher Updated APD batch assignment via API.
31. Jul 2019 13:26:23 CEST markus.moritz
23. Jun 2016 11:48:56 CEST Tobias Imported from Hamamatsu datasheet.