RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 1902020419 - Archived data from Tue, 11. August 2020 18:09:41 CEST


This is all the information about APD 1902020419. If it is wrong, edit the data.

Subdetector specification:  
Serial: 1902020419
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: B10
Break-through voltage: 403 V
Voltage for Gain 100 (T=+25°C): 375.4 V
Dark current: 18.2 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 209
Position in Box: 37
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 375.5021626 V     T = -25 °C: 340.2080355 V
Voltage for Gain 150: T = +20 °C: 383.2636278 V     T = -25 °C: 347.4996483 V
Voltage for Gain 200: T = +20 °C: 387.580081 V     T = -25 °C: 351.595456 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.477335434 V-1     T = -25 °C: 4.88470597 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.757922639 V-1     T = -25 °C: 9.092618854 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.5158853 V-1     T = -25 °C: 16.18330225 V-1
Break-through voltage: T = +20 °C: 395.8116373 V     T = -25 °C: 366.5272971 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:15:17 CET jreher Updated box and position via API.
11. Aug 2020 18:09:41 CEST jreher Updated APD characteristics via API.
21. Jul 2020 14:25:53 CEST jreher Updated APD location via API.
21. Jul 2020 14:25:43 CEST jreher Updated box and position via API.
29. Jun 2020 16:11:41 CEST jreher Updated APD batch assignment via API.
23. Jun 2016 11:48:48 CEST Tobias Imported from Hamamatsu datasheet.