RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 1902020400 - Archived data from Tue, 11. August 2020 18:09:36 CEST


This is all the information about APD 1902020400. If it is wrong, edit the data.

Subdetector specification:  
Serial: 1902020400
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: D13
Break-through voltage: 405 V
Voltage for Gain 100 (T=+25°C): 376.9 V
Dark current: 17 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 209
Position in Box: 18
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 376.6267829 V     T = -25 °C: 341.194811 V
Voltage for Gain 150: T = +20 °C: 384.5356653 V     T = -25 °C: 348.6027319 V
Voltage for Gain 200: T = +20 °C: 388.932539 V     T = -25 °C: 352.7812876 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.370696755 V-1     T = -25 °C: 4.860381362 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.266894423 V-1     T = -25 °C: 9.804734998 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.45004761 V-1     T = -25 °C: 15.43855472 V-1
Break-through voltage: T = +20 °C: 404.2254289 V     T = -25 °C: 367.9461973 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:15:15 CET jreher Updated box and position via API.
11. Aug 2020 18:09:36 CEST jreher Updated APD characteristics via API.
21. Jul 2020 14:25:51 CEST jreher Updated APD location via API.
21. Jul 2020 14:25:42 CEST jreher Updated box and position via API.
29. Jun 2020 16:11:15 CEST jreher Updated APD batch assignment via API.
23. Jun 2016 11:48:47 CEST Tobias Imported from Hamamatsu datasheet.