RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 1901020370 - Archived data from Tue, 11. August 2020 18:08:53 CEST


This is all the information about APD 1901020370. If it is wrong, edit the data.

Subdetector specification:  
Serial: 1901020370
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: B10
Break-through voltage: 412 V
Voltage for Gain 100 (T=+25°C): 383.5 V
Dark current: 13.1 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 330
Position in Box: 42
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 384.1810187 V     T = -25 °C: 348.2866038 V
Voltage for Gain 150: T = +20 °C: 392.0252757 V     T = -25 °C: 355.7825027 V
Voltage for Gain 200: T = +20 °C: 396.3946742 V     T = -25 °C: 359.9785454 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.524387205 V-1     T = -25 °C: 4.723395653 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.938182247 V-1     T = -25 °C: 9.517230407 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.62583548 V-1     T = -25 °C: 15.00049301 V-1
Break-through voltage: T = +20 °C: 411.9475069 V     T = -25 °C: 375.6010995 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:30:44 CET jreher Updated box and position via API.
11. Aug 2020 18:08:53 CEST jreher Updated APD characteristics via API.
21. Jul 2020 14:46:22 CEST jreher Updated APD location via API.
21. Jul 2020 14:46:13 CEST jreher Updated box and position via API.
29. Jun 2020 16:08:27 CEST jreher Updated APD batch assignment via API.
23. Jun 2016 11:48:46 CEST Tobias Imported from Hamamatsu datasheet.