RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 1802020074 - Archived data from Tue, 11. August 2020 18:10:16 CEST


This is all the information about APD 1802020074. If it is wrong, edit the data.

Subdetector specification:  
Serial: 1802020074
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: B06
Break-through voltage: 426 V
Voltage for Gain 100 (T=+25°C): 397.8 V
Dark current: 6.2 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 241
Position in Box: 14
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 398.2144676 V     T = -25 °C: 361.710236 V
Voltage for Gain 150: T = +20 °C: 406.1183639 V     T = -25 °C: 369.4504328 V
Voltage for Gain 200: T = +20 °C: 410.5090225 V     T = -25 °C: 373.7630481 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.501773991 V-1     T = -25 °C: 4.47838538 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.851501889 V-1     T = -25 °C: 8.708102919 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.55234496 V-1     T = -25 °C: 15.20301258 V-1
Break-through voltage: T = +20 °C: 425.8848276 V     T = -25 °C: 389.5277797 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:19:21 CET jreher Updated box and position via API.
11. Aug 2020 18:10:16 CEST jreher Updated APD characteristics via API.
21. Jul 2020 14:31:19 CEST jreher Updated APD location via API.
21. Jul 2020 14:31:09 CEST jreher Updated box and position via API.
29. Jun 2020 16:14:25 CEST jreher Updated APD batch assignment via API.
23. Jun 2016 11:47:39 CEST Tobias Imported from Hamamatsu datasheet.