RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 1716019602 - Archived data from Tue, 21. July 2020 14:37:32 CEST


This is all the information about APD 1716019602. If it is wrong, edit the data.

Subdetector specification:  
Serial: 1716019602
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: unknown
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: B07
Break-through voltage: 423 V
Voltage for Gain 100 (T=+25°C): 394.5 V
Dark current: 4.6 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 276
Position in Box: 47
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: none     T = -25 °C: none
Voltage for Gain 150: T = +20 °C: none     T = -25 °C: none
Voltage for Gain 200: T = +20 °C: none     T = -25 °C: none
Gain/Voltage slope at M = 100: T = +20 °C: none     T = -25 °C: none
Gain/Voltage slope at M = 150: T = +20 °C: none     T = -25 °C: none
Gain/Voltage slope at M = 200: T = +20 °C: none     T = -25 °C: none
Break-through voltage: T = +20 °C: none     T = -25 °C: none
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:23:58 CET jreher Updated box and position via API.
11. Aug 2020 17:51:56 CEST jreher Updated APD characteristics via API.
21. Jul 2020 14:37:42 CEST jreher Updated APD location via API.
21. Jul 2020 14:37:32 CEST jreher Updated box and position via API.
12. Dec 2019 12:01:05 CET jreher Updated box and position via API.
09. Dec 2019 14:52:56 CET jreher Updated box and position via API.
05. Dec 2019 18:02:09 CET spankonin Updated APD batch assignment via API.
23. Jun 2016 11:43:38 CEST Tobias Imported from Hamamatsu datasheet.