RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 1710019175 - Archived data from Tue, 11. August 2020 17:53:08 CEST


This is all the information about APD 1710019175. If it is wrong, edit the data.

Subdetector specification:  
Serial: 1710019175
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: G08
Break-through voltage: 425 V
Voltage for Gain 100 (T=+25°C): 395.8 V
Dark current: 4.1 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 283
Position in Box: 32
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 396.7578713 V     T = -25 °C: 360.1093486 V
Voltage for Gain 150: T = +20 °C: 404.6679814 V     T = -25 °C: 367.873976 V
Voltage for Gain 200: T = +20 °C: 409.0520608 V     T = -25 °C: 372.2266459 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.588844214 V-1     T = -25 °C: 4.630671851 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.229367456 V-1     T = -25 °C: 9.150794462 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.34940877 V-1     T = -25 °C: 14.15052427 V-1
Break-through voltage: T = +20 °C: 424.8923687 V     T = -25 °C: 388.3702141 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:24:51 CET jreher Updated box and position via API.
11. Aug 2020 17:53:08 CEST jreher Updated APD characteristics via API.
21. Jul 2020 14:39:24 CEST jreher Updated APD location via API.
21. Jul 2020 14:39:15 CEST jreher Updated box and position via API.
09. Dec 2019 18:29:54 CET spankonin Updated box and position via API.
09. Dec 2019 18:22:35 CET spankonin Updated APD batch assignment via API.
23. Jun 2016 11:42:29 CEST Tobias Imported from Hamamatsu datasheet.