RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 1709019103 - Archived data from Tue, 11. August 2020 17:52:56 CEST


This is all the information about APD 1709019103. If it is wrong, edit the data.

Subdetector specification:  
Serial: 1709019103
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: G07
Break-through voltage: 421 V
Voltage for Gain 100 (T=+25°C): 391.7 V
Dark current: 4.2 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 282
Position in Box: 22
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 392.1955674 V     T = -25 °C: 356.0560296 V
Voltage for Gain 150: T = +20 °C: 400.0766331 V     T = -25 °C: 363.7110813 V
Voltage for Gain 200: T = +20 °C: 404.4581325 V     T = -25 °C: 368.0029971 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.495847071 V-1     T = -25 °C: 4.726639478 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.857728621 V-1     T = -25 °C: 9.377819407 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.44813789 V-1     T = -25 °C: 14.61814788 V-1
Break-through voltage: T = +20 °C: 418.804525 V     T = -25 °C: 384.0736998 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:24:43 CET jreher Updated box and position via API.
11. Aug 2020 17:52:56 CEST jreher Updated APD characteristics via API.
21. Jul 2020 14:39:09 CEST jreher Updated APD location via API.
21. Jul 2020 14:39:00 CEST jreher Updated box and position via API.
09. Dec 2019 18:29:13 CET spankonin Updated box and position via API.
09. Dec 2019 18:23:50 CET spankonin Updated APD batch assignment via API.
23. Jun 2016 11:42:27 CEST Tobias Imported from Hamamatsu datasheet.