RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 1707018892 - Archived data from Tue, 06. October 2020 16:53:09 CEST


This is all the information about APD 1707018892. If it is wrong, edit the data.

Subdetector specification:  
Serial: 1707018892
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: G07
Break-through voltage: 413 V
Voltage for Gain 100 (T=+25°C): 384.2 V
Dark current: 3.3 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 258
Position in Box: 37
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 385.5088756 V     T = -25 °C: 349.5727716 V
Voltage for Gain 150: T = +20 °C: 393.3013842 V     T = -25 °C: 357.0500008 V
Voltage for Gain 200: T = +20 °C: 397.6237541 V     T = -25 °C: 361.2618042 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.468015207 V-1     T = -25 °C: 4.662101435 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.787946585 V-1     T = -25 °C: 9.301759563 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.36346106 V-1     T = -25 °C: 14.46888393 V-1
Break-through voltage: T = +20 °C: 405.9067555 V     T = -25 °C: 376.3142687 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:21:33 CET jreher Updated box and position via API.
08. Dec 2020 17:17:59 CET jreher Assigned to capsule by script
06. Oct 2020 16:53:09 CEST jreher Updated box and position via API.
11. Aug 2020 17:50:28 CEST jreher Updated APD characteristics via API.
21. Jul 2020 14:33:58 CEST jreher Updated box and position via API.
05. Dec 2019 13:58:47 CET jreher Updated box and position via API.
03. Dec 2019 14:46:42 CET jreher Updated APD location via API.
02. Dec 2019 19:19:01 CET spankonin Updated box and position via API.
02. Dec 2019 19:14:23 CET spankonin Updated APD batch assignment via API.
23. Jun 2016 11:42:22 CEST Tobias Imported from Hamamatsu datasheet.